TLP01-448-25x36

448 nm VersaChrome Edge™ tunable longpass filter

Product Code Description Lead Time Price Quantity Buy / Get Quote
TLP01-448-25x36 448 nm VersaChrome Edge™ tunable longpass filter 1-2 weeks £758.00
ADD
Click and drag to zoom the X-axis. For multi-line charts, click on the legend to show/hide lines.

Compare Charts

Specifications

Angle of Incidence:

0 – 60 °

Blocking Band:

ODavg > 6 348 – (0.98*Edge Wavelength 0deg) nm

Blocking Band, @ 60 deg:

ODavg > 6 348 – (0.975*Edge Wavelength 60deg)

Clear Aperture:

≥ 80% (elliptical)

Cone Half-angle:

Cuton Transition Width:

2.5% (from blocking to transmission)

Cuton Transition Width, @ 60 deg:

3% (from blocking to transmission)

Edge Wavelength, @ 0 Degree Angle:

minimum 448nm

Edge Wavelength, @ 60 Degree Angle:

maximum 400nm

Filter Effective Index:

1.76 Understanding 'Effective Index of Refraction' neff

Filter Thickness Tolerance, Unmounted:

± 0.1mm

Filter Thickness, Unmounted:

2.0mm

Mounted / Unmounted:

Unmounted

Optical Damage Rating:

Not tested

Orientation:

Reflective surface marked with part number – Orient in direction of incoming light

Product Type:

VersaChrome Edge Tunable Longpass Filter

Semrock Filter Family:

VersaChrome Edge™ Tunable Filters

Shape:

rectangular

Size:

25.2 x 35.6 x 2.0 mm

Substrate Thickness:

2.0mm

Substrate Type:

low-autofluorescence optical quality glass

Surface Quality (Scratch-Dig):

60-40

Transmission Band:

Tavg > 93% over 57 nm

Transmission Band, @ 60 deg:

Tavg > 90% over 51

Transverse Dimensions, LxW:

25.2 mm x 35.6 mm

Transverse Tolerance, Mounted:

± 0.1 mm

VersaChrome filters do what no thin-film filter has done before – offer tunability over a very wide range of wavelengths by adjusting the angle of incidence with essentially no change in spectral performance. They combine the spectral characteristics and two-dimensional imaging performance of a thin-film filter with the wavelength-tuning flexibility of a diffraction grating.

VersaChrome Edge tunable filters unlock virtually unlimited spectral flexibility for fluorescence microscopy and hyperspectral imaging, as well as for spectroscopy applications. By utilizing a combination of VersaChrome Edge tunable long-wave pass and short-wave pass filters, a bandpass filter as narrow as sub 5nm FWHM or as wide as 12% of the center wavelength throughout the visible and near-infrared wavelength ranges can be created. Semrock’s patented tunable thin-film filters can’t be found anywhere else in the market.